Merge Healthcare, Inc.: System, Image Processing, Radiological

Recall Summary Merge Healthcare's PACS medical imaging workstations have a software issue where thumbnail images from a previously viewed patient's study may incorrectly appear on screen while a different patient's study is open. This recall affects medical facilities nationwide in the US, as well as in Canada and the UK that use this imaging diagnostic system.

Recalling firm
Merge Healthcare, Inc.
Product
Merge PACS, medical imaging diagnostic workstation
Reason
Merge Healthcare has identified a potential issue that may cause thumbnails from a previously opened patient study to be displayed in the Thumbnail Bar while a different patient study is open.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-06-30
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, and UK.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/a0dbbd14-b6e6-44b6-8bdb-afd405289f00