Imris Imaging Inc: Ort300, Operating Room Table, Ort300 Modular Table Main Unit Assy Part Numbers 114093-000 And 114093-600
Recall Summary IMRIS Imaging Inc is recalling ORT300 operating room tables (Part Numbers 114093-000 and 114093-600) distributed worldwide because the locking mechanism may fail, preventing the table from properly locking or unlocking. This could cause unintended movement of the table during surgery and potentially injure patients. The recall affects hospitals and surgical facilities across all U.S. states and Canada, China, Japan, and Sweden.
- Recalling firm
- IMRIS Imaging Inc
- Product
- ORT300, operating room table, ORT300 Modular Table Main Unit Assy Part Numbers 114093-000 and 114093-600
- Reason
- The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- U.S. Food and Drug Administration
- Recall date
- 2026-08-06
- Status
- Ongoing
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, China, Japan, Sweden.