Imris Imaging Inc: Ort300, Operating Room Table, Ort300 Modular Table Main Unit Assy Part Numbers 114093-000 And 114093-600

Recall Summary IMRIS Imaging Inc is recalling ORT300 operating room tables (Part Numbers 114093-000 and 114093-600) distributed worldwide because the locking mechanism may fail, preventing the table from properly locking or unlocking. This could cause unintended movement of the table during surgery and potentially injure patients. The recall affects hospitals and surgical facilities across all U.S. states and Canada, China, Japan, and Sweden.

Recalling firm
IMRIS Imaging Inc
Product
ORT300, operating room table, ORT300 Modular Table Main Unit Assy Part Numbers 114093-000 and 114093-600
Reason
The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.
Classification
Class II
Category
medical device recall
Country
United States
Agency
U.S. Food and Drug Administration
Recall date
2026-08-06
Status
Ongoing
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, China, Japan, Sweden.