Vitek 2 Ast-ys08 Test Kits
- Recalling firm
- Medi'Ray New Zealand Ltd
- Product
- VITEK 2 AST-YS08 test kits. [Device]. Model: VITEK 2 AST-YS08 test kits manufactured on or after 18Jul2025.
- Reason
- Manufacturing Quality Control (QC) failures were observed during release testing for VITEK 2 AST-YS08 test kits manufactured on or after 18Jul2025. Quality Control MIC results for Amphotericin B were out-of-range-high (OORH). To be noted, there are no S/I/R interpretations for Amphotericn B. The antibiotic is reported as MIC only., If/When using AST-YS08 test kits manufactured on or after 18Jul2025, implement a VITEK 2 Systems BIOART rule to suppress analysis of the Amphotericin B antibiotic, and perform Amphotericin B testing via an alternate method if treatment with Amphotericin B is...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-03-04
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory