Vitek 2 Ast-ys08 Test Kits

Recalling firm
Medi'Ray New Zealand Ltd
Product
VITEK 2 AST-YS08 test kits. [Device]. Model: VITEK 2 AST-YS08 test kits manufactured on or after 18Jul2025.
Reason
Manufacturing Quality Control (QC) failures were observed during release testing for VITEK 2 AST-YS08 test kits manufactured on or after 18Jul2025. Quality Control MIC results for Amphotericin B were out-of-range-high (OORH). To be noted, there are no S/I/R interpretations for Amphotericn B. The antibiotic is reported as MIC only., If/When using AST-YS08 test kits manufactured on or after 18Jul2025, implement a VITEK 2 Systems BIOART rule to suppress analysis of the Amphotericin B antibiotic, and perform Amphotericin B testing via an alternate method if treatment with Amphotericin B is...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-03-04
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/9d461717-4079-4927-9b7c-517007ccd3af