Daavlin Distributing Company: Light, Ultraviolet, Dermatological

Recall Summary Daavlin Distributing Company is recalling Photera 400 phototherapy devices and ultraviolet lamps used for treating skin disorders because the UV output calibration is not calculating correctly—the devices are delivering less ultraviolet radiation than intended due to a sensor error. The recall affects devices distributed nationwide across all U.S. states and Washington, D.C. This is classified as a Class II recall, which indicates a serious issue though not one likely to cause severe harm.

Recalling firm
Daavlin Distributing Company
Product
Photera 400, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
Reason
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-03-05
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/9c4ebc1d-9642-4bb9-a160-6ccd27faf595