Ziehm-orthoscan Inc: Image-intensified Fluoroscopic X-ray System, Mobile

Recall Summary Ziehm-Orthoscan is recalling Orthoscan VERSA Accessory Carts and Versa Mini C-Arm Systems because welds on the cart may not be properly constructed, potentially causing the monitor arm to separate from the cart. The devices were distributed worldwide, including throughout the United States and several other countries. This is a Class II recall, meaning the defect could cause serious health consequences but is unlikely to result in death.

Recalling firm
ZIEHM-ORTHOSCAN INC
Product
Orthoscan VERSA Accessory Cart, REF: 1000-0021 and Versa Mini C-Arm System, REF: 1000-0020
Reason
The accessory cart weldment does not conform to weld dimensional specifications which may result in separation of the monitor arm mounting structure from the cart.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-09-03
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide - US Nationwide distribution in the states of AZ , CA , CO , CT , FL , GA , IL , IN , KS, MA , MD , MI , MN , NC , NE , NV , NY , OH , OR , SC, TX , UT , WA and the countries of Germany , Australia , Spain , Switzerland, Japan.