Recall Summary Medtronic is recalling the Octopus 4 Tissue Stabilizer (Model 29400), a heart surgery device, because the tubing was assembled incorrectly during manufacturing—the short tube was attached to the bottom and the long tube to the top, opposite of how it should be. This recall affects units distributed nationwide across all U.S. states. The device is used during heart surgery procedures to stabilize tissue during cardiac operations.

Recalling firm
Medtronic Perfusion Systems
Product
Octopus 4 Tissue Stabilizer, Model 29400
Reason
During the manufacturing process, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.
Classification
Class I
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-05-12
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide - US Nationwide distribution.

Official recall notice from FDA-DEVICES