Philips Ultrasound, Llc: System, Imaging, Pulsed Doppler, Ultrasonic
Recall Summary Philips Lumify Diagnostic Ultrasound Systems are recalled due to two potential problems: the system may fail to recognize previously registered transducers and require re-registration in areas without Wi-Fi or cellular access (such as ambulances or aircraft), and the connection between the transducer cable and the tablet/phone may become unstable or disconnect during use. These issues could delay diagnosis and treatment. The affected systems have been distributed nationwide in the U.S. and internationally.
- Recalling firm
- Philips Ultrasound, LLC
- Product
- Philips Lumify Diagnostic Ultrasound System
- Reason
- Ultrasound system may: 1) Not recognize previous transducer registration, prevent scanning, require re-registration with Wi-Fi/cellular access, which may be unavailable outside of the hospital (ambulances/aircraft/field); 2) At connection between transducer USB cable to tablet/phone, there may be an unstable connection/may disconnect disrupting function. May cause diagnosis/treatment delay.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- FDA-DEVICES
- Recall date
- 2026-09-24
- Status
- Open, Classified
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- Worldwide - US Nationwide distribution including in the states of CA, MA, IL, NY, OR, GA, TX, WA, FL, PA, WI, LA, NJ, KY, VA, MS, AZ, SC, CT, MN, UT, OH, AL, MI, NH, MO, TN, IN, DC, NE, NM, MT, NV, NC, MD, IA, SD, VT, KS, AK, CO, ID, OK, WV, AR, RI, ME, WY, HI, DE, ND and the countries of CA,...
Official recall notice from FDA-DEVICES
Recall details on Nibble Recalls: https://trynibble.app/recalls/99ddfadd-4f40-45e0-a6e7-79f73f14948c