Icu Medical, Inc.: Pump, Infusion

Recall Summary ICU Medical is recalling the Plum Solo Precision IV Pump (model 40001-0401) due to a software issue where the pump may fail to alert clinicians when medication delivery exceeds set limits. This could result in patients receiving incorrect doses of medication. The pump was distributed nationwide across the US and in Canada.

Recalling firm
ICU Medical, Inc.
Product
Plum Solo Precision IV Pump, 40001-0401
Reason
Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or over-delivery.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-04-29
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide - US Nationwide distribution in the states of ID, IL, IN, KS, MA, MD, NM, NY, PA, TN, TX and the country of Canada.

Official recall notice from FDA-DEVICES