Oscor Inc.: Introducer, Catheter

Recall Summary Oscor Inc. is recalling certain Impella heart pump introducer kits (14Fr and 23Fr sizes) distributed in Massachusetts due to manufacturing defects that could cause leaking from multiple points on the device. The leaking could occur from the sidearm, under the sheath cap, or along the hub area. This is a Class II recall affecting devices distributed in Massachusetts.

Recalling firm
Oscor Inc.
Product
Abiomed REF: 0052-3021, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 14F, 23F. heart pump and accessories.
Reason
Due to manufacturing issues, there is a potential for the 14Fr and 23Fr introducer sheaths to leak from the sidearm, under the sheath cap, and along the hub score lines in the hub region.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-06-10
Status
Open, Classified
Areas affected
MA
Distribution
Distribution in the US state of MA

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/9945aab9-8644-4509-8c56-e535eeb520c9