Raysearch Laboratories Ab: System, Planning, Radiation Therapy Treatment

RayCare 2024A SP1 Recall Summary RayCare 2024A SP1, a radiation therapy planning system used in cancer treatment, is being recalled because of software issues with treatment delivery notes that can result in data loss when notes are modified at the same time or updated during an active treatment session, with users not being alerted to these changes. The recall affects this oncology information system software distributed nationwide across all U.S. states and territories. This is classified as a Class II recall by RayCare's manufacturer, RaySearch Laboratories AB.

Recalling firm
RAYSEARCH LABORATORIES AB
Product
RayCare 2024A SP1. Oncology Information Systems.
Reason
Beam set delivery note issues in which there is potential data loss if the beam set delivery note is modified simultaneously and/or if a note is updated during a treatment session, the user is not notified of the change.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-05-20
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide - US Nationwide distribution.