Olympus - Insufflator, Endoscopic (japan)
- Recalling firm
- Olympus (NZ) Ltd
- Product
- Olympus - Insufflator, endoscopic (Japan). [Device]. Model: Unit models UHI, UHI-2, and UHI-3; Use available alternatives including an existing UHI-4 if available. Affected: All affected numbers -Olympus High Flow Insufflation Unit models - UHI, UHI-2, UHI-3 and N1000161. Also covers: UHI-4_High Flow Insufflator Unit
- Reason
- The manufacturer has determined that the software algorithm on High Flow Insufflation Unit, models UHI, UHI-2, and UHI-3, requires correction to address a potential issue which could lead to overpressure events. These events may occur due to an over insufflation of the abdominal cavity resulting from use of the UHI, UHI-2, or UHI-3 during patient procedures. This includes events where the device may not alarm or otherwise notify the user and/or may not relieve the over insufflation to the set pressure. Because no corrective solution is available, the manufacturer has decided to remove...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-01-26
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE
Recall details on Nibble Recalls: https://trynibble.app/recalls/981f2b0d-142e-4b60-82a1-a4a5852f3c0e