Dexcom Cgm System

Recalling firm
New Zealand Medical & Scientific Ltd
Product
Dexcom CGM system. [Device]. Model: SW12299 G7 App (Android) – Version 2.9.0 - 2.11.2 and SW12300 G7 App (iOS) – Version 2.9.0 - 2.11.2. Affected: Dexcom G7 App (iOS and Android), Versions 2.9-2.11.2, and the suppression of critical alerts.
Reason
The manufacturer has identied that when in 'Do Not Disturb' mode, critical alerts may not vibrate as expected and 'disk space critical,' 'recoverable error,' and 'Unrecoverable error' alerts may not sound or vibrate for up to 25 minutes. You're using an older version of the Dexcom G7 app, which may not perform as expected due to this issue., If a user is unaware of a prompt due to an alert not performing as expected, there is a potential for a missed / delayed detection of a hypoglycaemic or hyperglycaemic event.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-08-28
Status
Ongoing
Distribution
New Zealand - Retail

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/96f51697-aab8-4b19-aee0-cd9d5cef457e