Philips Medical Systems Nederland B.v.: Interventional Fluoroscopic X-ray System

Recall Summary Philips is recalling multiple models of its Azurion interventional fluoroscopic X-ray systems distributed nationwide because some units may contain unauthorized hard disk drives that could cause loss of imaging capability, loss of motorized movement, or loss of data. This is a Class II recall affecting Azurion 3, 5, and 7 systems with various model numbers. Healthcare facilities that use these systems should check if their equipment is included in the recall.

Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product
Philips Azurion under the following System Descriptions and corresponding model numbers: 1. Azurion 3 M12; Model Numbers: 722063, 722221. 2. Azurion 3 M15; Model Numbers: 722064, 722222, 722280. 3. Azurion 5 M12; Model Number: 722227. 4. Azurion 5 M20; Model Numbers: 722228, 722281. 5. Azurion 7 B12; Model Numbers: 722067, 722225. 6. Azurion 7 B20; Model Numbers: 722068, 722226. 7. Azurion 7 M12; Model Numbers: 722078, 722223. 8. Azurion 7 M20; Model Numbers: 722079, 722224, 722282. Interventional Fluoroscopic X-Ray System.
Reason
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-06-17
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution.