Diagnostica Stago, Inc.: Test, Qualitative And Quantitative Factor Deficiency

Recalling firm
Diagnostica Stago, Inc.
Product
STA Staclot Protein S, Model/Catalog Number: 00746-Reagent Lot number 273074; qualitative and quantitative factor deficiency test
Reason
A positive drift in QC values within the reagent's stated 4-hour onboard stability period was identified
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-07-29
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution in the states and territory of NJ, NY, NJ, VA, PA, FL, AL, GA, TN, OH, IL, MO, NE, TX, CO, UT, CA, PR.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/95fdad92-1e9d-45a1-9f41-e5f3d2ea7022