Siemens Medical Solutions Usa, Inc: Interventional Fluoroscopic X-ray System
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray system
- Reason
- During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- FDA-DEVICES
- Recall date
- 2026-03-09
- Status
- Open, Classified
- Distribution
- US and Worldwide