Philips North America: Philips Ingenia 3.0t Cx With Mr Elastography (mre). 1. Model Number (ref): 781271. 2. Model Number (ref): 782105. 510(k) Numbers: K173079, K183063,...

Recall Summary Philips is recalling certain Ingenia 3.0T CX MRI machines (models 781271 and 782105) worldwide because specific imaging settings can cause measurement errors in elastography scans, resulting in incorrect voxel size calculations. These affected MRI machines are used in medical facilities across the US and numerous countries for diagnostic imaging. This is classified as a Class II recall, indicating a potential for serious health consequences.

Recalling firm
Philips North America
Product
Philips Ingenia 3.0T CX with MR Elastography (MRE). 1. Model Number (REF): 781271. 2. Model Number (REF): 782105. 510(k) Numbers: K173079, K183063, K193215, K213583, K223458, K251397, K251808.
Reason
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
Classification
Class II
Category
medical device recall
Country
United States
Agency
U.S. Food and Drug Administration
Recall date
2026-04-14
Status
Ongoing
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, Cuba, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Lebanon, Lithuania, Malta, Mexico, Myanmar,...

Official recall notice from U.S. Food and Drug Administration

Recall details on Nibble Recalls: https://trynibble.app/recalls/94215145-ffef-483a-bc0e-b996eecff83c