Noxboxi Nitric Oxide Monitor And Delivery System
- Recalling firm
- Keyport Ltd
- Product
- NOXBOXi Nitric Oxide Monitor and Delivery System. [Device]. Model: Product Code - NOXBOX-I. Affected: Serial number: NI103206, NI100931, NI101157, NI101195, NI103009, NI100792
- Reason
- The manufacturer has identified a manufacturing issue affecting certain oxygen sensors. The issue may cause oxygen sensors to intermittently display irregular oxygen concentrations or cause the NOxBOXi device to fail the required oxygen sensor zero calibration. This sensor failure may be, intermittent and may not be detected immediately following manufacture or during pre-use system and setup checks. Based on the design of the NOxBOXi device, the oxygen sensor is used solely to monitor the oxygen level in the sampled gas and is complimentary to other oxygen monitoring that might be used on...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-10-22
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE
Recall details on Nibble Recalls: https://trynibble.app/recalls/941b229d-1537-4f46-a225-619586d19164