Angiographic X-ray

Azurion Recall Summary Philips Medical Systems Nederland BV is recalling Azurion products distributed to hospitals, pharmacies, and laboratories in New Zealand. The manufacturer will be issuing guidance on how to use these products. This is classified as a Class I recall, the most serious type.

Recalling firm
Philips New Zealand Commercial Limited
Product
Angiographic X-Ray. [Device]. Model: Philips Azurion and Allura Systems; Philips Azurion. Also covers: Azurion
Reason
The manufacturer has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch due to:, · Intermittent or absent electrical contact when the foot switch pedal is pressed due to gradual formation of silicon oxide layers on the microswitch1 contact surfaces., · Internal damage to the foot switch cable., · Application of strain relief2 on the connector that may not fully mitigate mechanical load, which could potentially result in connector damage and/or intermittent or complete loss of connection.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-03-26
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/93e23d56-e41a-44aa-8a1b-ab9c4d09a70b