Aniara Diagnostica Llc: Calibrator, Secondary

Recall Summary Aniara Diagnostica LLC is recalling the 5-PLASMA Bivalirudin AMR Verification Kit (Model 5D-45905) because the reconstituted plasma and bivalirudin material degrades over time, causing inaccurate measurements of bivalirudin levels in the controls and calibrators. This is a Class II recall affecting medical laboratories nationwide that received this product. The company distributed the kits primarily in Illinois, North Carolina, and Tennessee, though they were also sent to other states.

Recalling firm
Aniara Diagnostica LLC
Product
5-PLASMA Bivalirudin AMR Verification Kit, Model Number 5D-45905
Reason
Stability issues with the reconstituted lyophilized human plasma and bivalirudin matrix have been identified. These stability issues lead to a degradation of measurable bivalirudin in the controls, calibrators, and linearity values.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-07-01
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution in the states of IL, NC, TN.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/8f203173-378a-457d-9097-64a6c939a668