Accuray Cyberknife Treatment Delivery System

Recalling firm
Clinical & Regulatory Services Limited
Product
Accuray CyberKnife Treatment Delivery System. [Device]. Model: CyberKnife Treatment Delivery System Equipped with the Xchange Robotic Collimator Changer
Reason
The manufacturer identified a potential problem during CyberKnife® collimator exchange. In certain cases, during a housing exchange operation, a collimator housing may not fully latch or release. This event triggers a system interlock while over the Xchange® table that prevents motion from the operator console. In this state, the collimator housing may fall from the CyberKnife robot while over the Xchange table. In normal use, patients should be removed from the treatment room following this system interlock. While motion is prevented from the operator console, users are capable of...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-06-12
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/8d79ab22-ef2c-433e-954b-13d9ddc73dd4