Accuray Cyberknife Treatment Delivery System
- Recalling firm
- Clinical & Regulatory Services Limited
- Product
- Accuray CyberKnife Treatment Delivery System. [Device]. Model: CyberKnife Treatment Delivery System Equipped with the Xchange Robotic Collimator Changer
- Reason
- The manufacturer identified a potential problem during CyberKnife® collimator exchange. In certain cases, during a housing exchange operation, a collimator housing may not fully latch or release. This event triggers a system interlock while over the Xchange® table that prevents motion from the operator console. In this state, the collimator housing may fall from the CyberKnife robot while over the Xchange table. In normal use, patients should be removed from the treatment room following this system interlock. While motion is prevented from the operator console, users are capable of...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-06-12
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE
Recall details on Nibble Recalls: https://trynibble.app/recalls/8d79ab22-ef2c-433e-954b-13d9ddc73dd4