Trinity Biotech Premier Hb9210™ Hba1c Analyser: Risk Of Positive Bias And Updates To Instructions For Use (ifu), Including Use As A Diagnostic Aid In Diabetes Mellitus (dsi/2025/003)

Recall Summary The Trinity Biotech Premier Hb9210 HbA1c Analyser, a device used to measure blood sugar levels for diabetes diagnosis, is being recalled because it produces results that are consistently higher than they should be. This has led to patients being incorrectly diagnosed with pre-diabetes or diabetes when they may not actually have these conditions. The device is distributed in the United Kingdom.

Recalling firm
Risk of Positive Bias and Updates to Instructions for Use (IFU)
Reason
The MHRA has received reports describing a positive bias in HbA1c results delivered by the Trinity Biotech Premier Hb9210 HbA1c analyser, which has resulted in patients being incorrectly diagnosed as pre-diabetic or diabetic.
Classification
Unclassified
Category
drug recall
Country
United Kingdom
Agency
UK Medicines and Healthcare products Regulatory Agency
Recall date
2025-07-22
Status
Ongoing
Distribution
United Kingdom

Official recall notice from UK Medicines and Healthcare products Regulatory Agency

Recall details on Nibble Recalls: https://trynibble.app/recalls/8d4fc8c4-8732-426d-9cc9-507e0eda9a96