Baxter Healthcare Corporation: Table, Operating-room, Electrical
- Recalling firm
- Baxter Healthcare Corporation
- Product
- TruSystem 7000 U Operating Table (dV), Product Code 1723633
- Reason
- The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.
- Category
- medical device recall
- Country
- United States
- Agency
- FDA-DEVICES
- Recall date
- 2026-09-14
- Status
- Ongoing
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- US Nationwide distribution in the states of IL, MO, TX.
Official recall notice from FDA-DEVICES
Recall details on Nibble Recalls: https://trynibble.app/recalls/8cee8ffa-67f7-49d2-a146-d10c00deb769