Heraeus Medical Gmbh (dental Division) Philipp-reis-str. 8 Wehrheim Germany: Mixer, Cement, For Clinical Use

Recalling firm
Heraeus Medical GmbH (Dental Division) Philipp-Reis-Str. 8 Wehrheim Germany
Product
Heraeus, PALAMIX uno. Material Number: 66057893.
Reason
Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).
Classification
Class I
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-03-25
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide distribution - US Nationwide and the countries of Aland Islands, Algeria, Australia, Austria, Belgium, Brazil , Chile, China, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Kosovo, Lithuania, Luxembourg, Malaysia,...

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/8cd91a46-1de3-4775-b653-8f0683ca61b3