Infusion Pump Administration Set

Recalling firm
Becton Dickinson Ltd
Product
Infusion pump administration set. [Device]. Model: Alaris™ Products. Alaris™ VP Volumetric Pump. Alaris™ Safety Clamp. INFUSION SET. 15 µm FILTER. Affected: Product code - 70593 lot number 1032633
Reason
The manufacturer has identified, as a result of customer complaints received, that the specific product code and lot number of Alaris™ VP Volumetric Pump Infusion Sets have the potential for fluid leakage from the silicone tube at the end of the pumping segment., Leakage at the silicone pumping segment during priming or infusion may result in under-infusion of prescribed therapy, unintended exposure to hazardous drugs, and increased risk of air embolism or bacterial contamination if urgent set replacement compromises sterility.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-01-27
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE