Infusion Pump Administration Set
- Recalling firm
- Becton Dickinson Ltd
- Product
- Infusion pump administration set. [Device]. Model: Alaris™ Products. Alaris™ VP Volumetric Pump. Alaris™ Safety Clamp. INFUSION SET. 15 µm FILTER. Affected: Product code - 70593 lot number 1032633
- Reason
- The manufacturer has identified, as a result of customer complaints received, that the specific product code and lot number of Alaris™ VP Volumetric Pump Infusion Sets have the potential for fluid leakage from the silicone tube at the end of the pumping segment., Leakage at the silicone pumping segment during priming or infusion may result in under-infusion of prescribed therapy, unintended exposure to hazardous drugs, and increased risk of air embolism or bacterial contamination if urgent set replacement compromises sterility.
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-01-27
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE