Infinity Ammonia Standard Set

Recalling firm
Thermo Fisher Scientific New Zealand Ltd
Product
Infinity Ammonia Standard Set. [Device]. Model: Catalog number: TR60901 - Lot number: 382963, 364987, 307166
Reason
The manufacturer has identified that elevated levels of ammonia may be either: (a) due to inborn errors of metabolism; or (b) secondary to other conditions, such as viral hepatitis or cirrhosis. This Standard is not a companion diagnostic, nor is it intended for use in a specific population. This reagent is intended to be an aid to diagnosis, for laboratory professional use only., A complaint was received from a customer reporting approximately 6% to 15% low calibration bias after switching ammonia standard lots and attributed the bias to the calibrators., This issue may lead to prognostic...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-11-02
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/8c627164-3fb1-4b8d-8b03-9d67614c0ba6