Olympus - Sheaths Laser Sidefire

Recalling firm
Olympus (NZ) Ltd
Product
Olympus - Sheaths Laser Sidefire. [Device]. Model: SHEATH LASER SIDEFIRE OUTER 22.5Fr. Affected: All serial numbers
Reason
In October 2024, Olympus issued a Product Correction in relation to the Cystoscope Outer sheath, 22.5 Fr (Medsafe Ref #33877). The Product Correction Notice instructed users that the Cystoscope Outer sheath should not be used with a GreenLight Laser for BPH therapy until further notice. The manufacturer has received reports about a Cystoscope Outer sheath damaged tip during use of a laser probe. The manufacturer conducted additional testing, which indicated further regulatory actions would be required in order to state compatibility of the cystoscopy sheath with GreenLight laser fibres....
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-01-13
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory