Natus Medical Incorporated: Ventilatory Effort Recorder

Recalling firm
NATUS MEDICAL INCORPORATED
Product
Embletta MPR PG Multi Parameter Recorder Sleep Monitor, Model/Catalog Number: 2003011
Reason
The device may be unable to correctly format its internal memory. When the internal memory cannot be correctly formatted, previously recorded studies may remain on the device and may be updated with incorrect patient information from a subsequent study. The issue may also prevent studies from initiating recording successfully or may cause recordings to end prematurely, resulting in incomplete or unavailable diagnostic data and the potential need to repeat a patient study.
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-09-29
Status
Ongoing
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, China, Italy, Portugal.

Official recall notice from FDA-DEVICES