Physio-control Lifepak 35 Defibrillator/monitor

Recall Summary Physio-Control LIFEPAK 35 defibrillator/monitor devices are being recalled due to two manufacturing issues. First, under certain manual mode settings, the device may fail to display a "CHECK RHYTHM / ECG Analysis Inconclusive" message, though the device still correctly analyzes rhythms and delivers therapy as needed. Second, some printer and modem connectors may not be properly seated in the rear housing. These devices were distributed to hospitals, pharmacies, and laboratories in New Zealand.

Recalling firm
Stryker New Zealand Ltd
Product
Physio-Control LIFEPAK 35 defibrillator/monitor. [Device]. Model: LIFEPAK 35 Monitor/Defib. Affected: Catalogue number - 70335-000016, 70335-000017, 70335-000030, 70335-000018, 70335-000034
Reason
The manufacturer has identified a problem with the LP35 where the device does not display the message "CHECK RHYTHM / ECG Analysis Inconclusive" as intended when all below conditions are present simultaneously. The problem does not affect the display of "Shock Advised" or "No Shock Advised" messages, and therapy delivery remains available. Additionally, this nonconformance is limited to the device on-screen prompts only and rhythm analysis is not impacted by this problem., Problem 1: Under a specific combination of non-default Manual Mode settings, the "CHECK RHYTHM / ECG Analysis...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-09-14
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE