Aisys Anaesthesia Work Stations
- Recalling firm
- GE Healthcare Limited
- Product
- Aisys Anaesthesia Work Stations. [Device]. Model: Aisys CS2, Aisys CS2 with Et Control and S/5 Aisys Anesthesia Delivery Systems. Affected: Replaceable parts: 1011-8000-000-S - Gas Mixer Electronic – Assembly. 5700665 - R-FMI34141 GAS MIXER ELECTRONIC
- Reason
- The manufacturer has become aware that Aisys CS2, Aisys CS2 with Et Control and S/5 Aisys Anesthesia Delivery Systems, containing certain electronic gas mixer assemblies, can automatically transition to Alternate O2 control., In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms ("Set Alt O2 Flow! Agent delivery off!" or "Set Alt O2 flow! Check agent setting!")., When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-09-23
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE