Philips North America: System, Nuclear Magnetic Resonance Imaging

Recall Summary Philips is recalling its SmartPath to dStream 3.0T MRI system with MR Elastography (model 782145) because certain image reconstruction settings can produce incorrect stiffness measurements when used with Resoundant's algorithm. The issue affects units distributed worldwide, including all U.S. states and numerous countries. This is classified as a Class II recall, meaning the problem could cause adverse health effects but is not life-threatening.

Recalling firm
Philips North America
Product
Philips SmartPath to dStream for 3.0T with MR Elastography (MRE). 1. Model Number (REF): 782145.
Reason
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-04-14
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, Cuba, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Lebanon, Lithuania, Malta, Mexico, Myanmar,...

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/87790323-2307-4c78-8268-dbb1243f1c5f