Noxboxi Nitric Oxide Monitor And Delivery System

Recalling firm
Keyport Ltd
Product
NOXBOXi Nitric Oxide Monitor and Delivery System. [Device]. Model: NOxBOXi Intelligent Delivery and Monitor. Affected: Product code: NOXBOX-1
Reason
The manufacturer has identified an increased trend in complaints for oxygen leaks within the NOxMixer module of the NOxBOXi device due to failure of the tubing or tubing connection. The NOxMixer module is intended to deliver a continuous flow of nitric oxide (NO), mixed in line with oxygen (O2) for emergency applications and when manually ventilating a patient (i.e., during Emergency and Manual Bagging modes). This failure occurs inside the NOxMixer module when a leak develops in the internal oxygen tubing or the tubing connection at the elbow fitting upstream of the oxygen flowmeter.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-08-31
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/8737c5ab-cbbf-4668-b22e-3a22235108a8