Zimmer, Inc.: Prosthesis, Knee, Patello/femorotibial, Semi-constrained, Uncemented, Porous, Coated, Polymer/metal/polymer

Recall Summary Zimmer, Inc. is recalling certain Persona Revision Trabecular Metal Femoral Distal Augments (Size 9, 9+, and 10mm thickness) used in knee replacement surgery. A posterior augment was found packaged inside a distal augment package instead of in its correct location. No patient injuries have been reported from this packaging error.

Recalling firm
Zimmer, Inc.
Product
Brand Name: Persona Revision Product Name: Persona Revision Trabecular Metal Femoral Distal Augment Model/Catalog Number: 42-5572-066-10 Software Version: N/A Product Description: Persona Revision Trabecular Metal Femoral Distal Augment Size 9, 9+ 10mm Thicknes Component: N/A
Reason
Firm received a complaint with no clinical impact in which a posterior augment was found within the distal augment package.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-05-11
Status
Open, Classified
Distribution
Refer to the attached ZFA 2026-00088_Initial Consignee List and ZFA 2026-00088_Initial Distribution History containing the distribution and the identification of the affected consignees.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/87263ba3-bbd2-4016-a266-78d7e919880e