[mdr]philips Radiography 7000 M System
- Recalling firm
- Philips New Zealand Commercial Limited
- Product
- [MDR]Philips Radiography 7000 M system. [Device]. Model: Philips Radiography 7000 M system (Model 712010)
- Reason
- The manufacturer has become aware of a potential safety issue impacting the Radiography 7000 M system (Model 712010). The issue may occur with the Radiography 7000 M system immediately after an exposure, causing the drive function to fail and preventing repositioning of the system. The system may remain stationary instead of driving upon activation of the handlebar. The issue is intermittent and has been observed during routine mobile radiography use.
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-10-08
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE
Recall details on Nibble Recalls: https://trynibble.app/recalls/833b67fa-e1b7-4d32-bb95-3c7e68e9eea8