[mdr] Radiography 7000 M
- Recalling firm
- Philips New Zealand Commercial Limited
- Product
- [MDR] Radiography 7000 M. [Device]. Model: Radiography 7000 M - REF number: 712010
- Reason
- The company had been aware that when the Radiography 7000 M system is powered ON and connected to facility electrical mains for charging, it is possible the facility circuit breaker may be interrupted and prevent system charging. This problem may occur when facilities have Residual Current Detection (RCD) installed on electrical mains supply and have 30mA circuit breaker., This problem occurs during charging and not during clinical use.
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-06-26
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE
Recall details on Nibble Recalls: https://trynibble.app/recalls/831a8026-feac-477d-ae63-90d888b66bab