[mdr] Radiography 7000 M

Recalling firm
Philips New Zealand Commercial Limited
Product
[MDR] Radiography 7000 M. [Device]. Model: Radiography 7000 M - REF number: 712010
Reason
The company had been aware that when the Radiography 7000 M system is powered ON and connected to facility electrical mains for charging, it is possible the facility circuit breaker may be interrupted and prevent system charging. This problem may occur when facilities have Residual Current Detection (RCD) installed on electrical mains supply and have 30mA circuit breaker., This problem occurs during charging and not during clinical use.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-06-26
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/831a8026-feac-477d-ae63-90d888b66bab