Medline Industries, Lp: Cardiac Catheterization Kit

Recall Summary Medline is recalling 14 different cardiac procedure kits used in heart surgeries and related procedures because some samples failed quality tests. The problems included rough surfaces on catheters, connectors that detached too easily, and catheters with incorrect outer diameters. These kits were distributed nationwide in the US and to several other countries.

Recalling firm
Medline Industries, LP
Product
MEDLINE Medical Procedure Kits labeled as: 1) BASIC HEART PACK, Medline Kit/SKU DYNJ67331B; 2) CARDIAC SURGERY, Medline Kit/SKU DYNJ911589A; 3) CAROTID ENDARTECOMY, Medline Kit/SKU DYNJ36211D; 4) CHLA ECMO / CARDIAC PACK, Medline Kit/SKU DYNJ87973; 5) CHLA ECMO / CARDIAC PACK, Medline Kit/SKU DYNJ87973A; 6) CV PART 1 AND 2, Medline Kit/SKU DYNJ908879; 7) CV PART 1 AND 2, Medline Kit/SKU DYNJ908879C; 8) FEM POP CDS, Medline Kit/SKU CDS983670K; 9) LAP BASIN TRAY, Medline Kit/SKU DYNJ50908B; 10) OPEN HEART LTX CATHETERS KIT, Medline Kit/SKU DYNJ0049032A; 11) OPEN HEART PACK, Medline Kit/SKU...
Reason
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-05-04
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/8197233f-b15d-4db8-9311-c8dc9b567289