Philips Tempus Pro Patient Monitor, Ref: 00-1004, 00-1007, 00-1004-r, 00-1007-r, 00-1024-r, 00-1026-r. USB 2-channel Ibp Module, Ref: 01-2017. Used With User/operator Manual

Recall Summary Philips Tempus Pro Patient Monitors (multiple model numbers) and USB 2-Channel IBP Modules are being recalled because their measurements of intracranial pressure and bladder pressure have not been tested or validated for accuracy. Healthcare providers using these devices may make incorrect clinical decisions if they rely on these unvalidated measurements. This recall affects medical facilities using these monitoring systems.

Recalling firm
Remote Diagnostic Technologies Ltd. Farnborough Aerospace Centre Aerospace Boulevard Farnborough United Kingdom
Product
Philips Tempus Pro Patient Monitor, REF: 00-1004, 00-1007, 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R. USB 2-Channel IBP Module, REF: 01-2017. Used with User/Operator Manual
Reason
Patient monitor measures Intracranial Pressure (ICP) and Bladder Pressure (BDR) but ICP and BDR measurements have not been tested or validated for accuracy. If healthcare providers base clinical judgment on unvalidated measurements, this could potentially lead to incorrect clinical decisions.
Classification
Class II
Category
consumer product recall
Country
United States
Agency
FDA-DEVICES
Recall date
2025-11-26
Status
Open, Classified

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/816f3c7e-0601-4bbf-b4b8-d9501df9629d