Philips Tempus Pro Patient Monitor, Ref: 00-1004, 00-1007, 00-1004-r, 00-1007-r, 00-1024-r, 00-1026-r. USB 2-channel Ibp Module, Ref: 01-2017. Used With User/operator Manual
Recall Summary Philips Tempus Pro Patient Monitors (multiple model numbers) and USB 2-Channel IBP Modules are being recalled because their measurements of intracranial pressure and bladder pressure have not been tested or validated for accuracy. Healthcare providers using these devices may make incorrect clinical decisions if they rely on these unvalidated measurements. This recall affects medical facilities using these monitoring systems.
- Recalling firm
- Remote Diagnostic Technologies Ltd. Farnborough Aerospace Centre Aerospace Boulevard Farnborough United Kingdom
- Product
- Philips Tempus Pro Patient Monitor, REF: 00-1004, 00-1007, 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R. USB 2-Channel IBP Module, REF: 01-2017. Used with User/Operator Manual
- Reason
- Patient monitor measures Intracranial Pressure (ICP) and Bladder Pressure (BDR) but ICP and BDR measurements have not been tested or validated for accuracy. If healthcare providers base clinical judgment on unvalidated measurements, this could potentially lead to incorrect clinical decisions.
- Classification
- Class II
- Category
- consumer product recall
- Country
- United States
- Agency
- FDA-DEVICES
- Recall date
- 2025-11-26
- Status
- Open, Classified
Official recall notice from FDA-DEVICES
Recall details on Nibble Recalls: https://trynibble.app/recalls/816f3c7e-0601-4bbf-b4b8-d9501df9629d