Liquichek Cardiac Markers Plus Control Lt
Bio-Rad is recalling Liquichek Cardiac Markers Plus Control LT, a quality control product used in hospitals, pharmacies, and laboratories in New Zealand to verify the accuracy of cardiac marker tests. The instructions for use need to be updated. This is classified as a Class I recall, the most serious category.
- Recalling firm
- Bio-Rad Laboratories Inc
- Product
- Liquichek Cardiac Markers Plus Control LT. [Device]. Affected: Affected lot numbers - 1003150 series, 1003160 series
- Reason
- Internal QC testing reported that BNP and proBNP in-use (thawed opened) stability claims for Liquichek Cardiac Markers Plus LT Control Lot 1003150 & 1003160 were not met. The stability claims have been revised from 5 days to 3 days. All other claims for this lot of Liquichek Hematology Control (C) remain unchanged.
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-05-27
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE