Crhf - Carelink Smartsync Device Manager System

Recalling firm
Medtronic New Zealand Limited
Product
CRHF - CareLink SmartSync Device Manager System. [Device]. Model: Software Update Available: CareLink™ SmartSync™ Device Manager; CareLink™ SmartSync™ Micra VR2/AV2 Application software D00U022. Also covers: CRHF - Virtual Express Patient Connector
Reason
A software update (version 4.13.2 or higher) for the CareLink™ SmartSync™ Device Manager (SmartSync) is now available. This SmartSync update eliminates the potential for an inaccurately displayed longevity estimate on Micra VR2/AV2 software screens. Device performance, including battery voltage and longevity, remain unaffected.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-09-25
Status
Ongoing
Distribution
New Zealand - Healthcare Professional

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/8138307e-863c-47a9-9e1e-32b74f3c64db