Medtronic Minimed, Inc.: Calculator, Drug Dose

Recalling firm
Medtronic MiniMed, Inc.
Product
Generic Name: MiniMed Go App Generic Name: MiniMed Go App Model/Catalog Number: MMT-8600 (iOS), MMT-8601 (Android) Software Version: Software version 1.2.1 Product Description: The MiniMed Go app for iOS or MiniMed Go app for Android is compatible with the following components: " Simplera sensor " Instinct sensor " InPen smart insulin pen " Apple Watch [Optional] Continuous glucose monitoring (CGM) is a technology that allows you to continuously view glucose values. The app uses an inserted Simplera sensor CGM or Instinct sensor CGM with Bluetooth wireless technology that converts small...
Reason
A software defect in which existing users who had previously configured and enabled the long-acting dose reminder feature in version 1.2.0 or earlier experienced failure of the long-acting dose reminder after updating their app to version 1.2.1. The long-acting dose reminder is an optional feature that, when enabled, triggers a notification at the scheduled time to remind users to take their long-acting insulin dose. The issue was caused by a type cast exception during database migration that occurs upon app update to version 1.2.1 causing a silent failure that prevented the creation of...
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-09-02
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide - U.S. Nationwide distribution in the states of AL, CA, CT, FL, LA, MA, MD, NY, OH, PA, and TX. The countries of Austria, Belgium, Denmark, Finland, Germany, Iceland, Italy, Norway, Spain, Sweden, Switzerland, United Kingdom.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/807e5151-ed7c-4a94-b9f2-d6aa36e6ece5