Vitros Chemistry Products Mg Slides
- Recalling firm
- Ortho-Clinical Diagnostics
- Product
- VITROS Chemistry Products Mg Slides. [Device]. Model: Catalog number - 825 5093 - 60-Slide Cartridges 192 1204 - 18-Slide Cartridges. Affected: All in-date and future lots
- Reason
- The manufacturer became aware of the problem through customer complaints. The VITROS Mg Slides Instructions For Use (IFU) states an On-analyzer stability of = 2 weeks, when the VITROS System is turned on; however, investigation has determined that VITROS Mg Slides results may experience a negative drift outside acceptable limits (greater than 2 within-lab Standard Deviations (WLSD)) sooner than 2 weeks, when cartridges are stored on the VITROS System. This problem affects all current in-date and future lots of VITROS Mg Slides.
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-03-02
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE