Biofire Blood Culture Identification 2 (bcid2) Panel

Recalling firm
Medi'Ray New Zealand Ltd
Product
BIOFIRE Blood Culture Identification 2 (BCID2) Panel. [Device]. Affected: Lots# 0004102964, 0004102996, 0004102946, 0004102956, and 0004102998 - All unexpired lots when used with specific BACT/ALERT® Blood Culture Bottles.
Reason
The manufacturer has identified an increased risk of false positive Serratia marcescens results when the BIOFIRE BCID2 Panel is used with certain lots of BACT/ALERT Culture Bottles., The presence of non-viable organism does not compromise the intended function of the blood culture bottles (culturing viable microorganisms). However, the BIOFIRE BCID2 Panel detects nucleic acid from viable and non-viable organisms alike.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-03-31
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE