Hexavue Focus Integration System
- Recalling firm
- Steris New Zealand Limited
- Product
- HexaVue Focus Integration System. [Device]. Model: HexaVue Focus system that is running software version 2.0.18.
- Reason
- The manufacturer identified through customer complaints that a software issue may result in unintended mismatch of patient information on manually printed images. Specifically, during an active procedure for Patient A, a user may manually select a different patient record (Patient B) from the �Patient Cases� menu and subsequently manually print images. In this specific manual print workflow, the printed output may display patient identifying information from the selected patient record (Patient B) rather than the active procedure (Patient A), including patient name, date of birth, and...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-09-15
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE
Recall details on Nibble Recalls: https://trynibble.app/recalls/7da2d879-2cfc-4fc6-abf7-1864f8f26411