Iq200 Series Analyzers And Dxu 840m/850m Iris Analyzers
- Recalling firm
- Beckman Coulter NZ Ltd
- Product
- iQ200 Series Analyzers and DxU 840m/850m Iris Analyzers. [Device]. Model: iQ200 Series Analyzers and DxU 840m/850m Iris Analyzers. Affected: All serial numbers
- Reason
- The manufacturer has received customer complaints about lamina fluid leaks in iQ200 Series Analyzers and DxU 840m/850m Iris Analyzers. Investigation determined that the leaks are caused by a check valve in the rinse line becoming structurally compromised. This may allow lamina fluid to migrate to internal instrument surfaces and eventually leak onto the table surface and/or floor.
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-08-12
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory