Flexicare - Non-rebreathing Mask

Recalling firm
Jackson Allison Medical and Surgical Ltd
Product
Flexicare - Non-Rebreathing Mask. [Device]. Model: 032-10-075NCNZ Flexicare Adult Non-Rebreathing Green Tinted Nose Clip Mask Head Strap Box/50 (26219001080). Affected: All Lot numbers
Reason
The affected device has been incorrectly labelled as a non-rebreathing mask. In practice, the mask functions as a rebreathing mask, as it does not include the required visible one-way valves over the side exhalation ports. These valves are essential components that prevent re-inhalation of exhaled gases in a true non-rebreathing system.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-03-03
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/7d52c702-f72a-4fe7-8a33-c8c2b654d154