Bd Connecta Stopcock With Extension Tube

Recalling firm
Becton Dickinson Ltd
Product
BD Connecta Stopcock with Extension Tube. [Device]. Model: Product code: 394995 BD Connecta™ Stopcock with Extension Tube - white 16 cm. Affected: Lot number: 5042456, 5062802, 5073966, 5073969, 5098804, 5120823
Reason
The manufacturer has identified that specific lot numbers 394995 of the BD Connecta Stopcock with Extension Tube product may have been manufactured with an inadequate amount of solvent which could result in separation of the tubing from the male luer component., Separation of the Connecta™ Stopcock with Extension Tube may occur during use, resulting in a compromised fluid pathway. This may result in leakage with the potential consequences including under-infusion or disruption of prescribed therapy, exposure to potentially hazardous fluids, infection or air embolism., The separation may be...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-11-18
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/7d0c192b-7719-4944-ad8b-7861b696a6bc