Biofire Filmarray Torch System Base

Recalling firm
Medi'Ray New Zealand Ltd
Product
BIOFIRE FILMARRAY TORCH System Base. [Device]. Model: Ref number - HTFA-ASY 0104, Shipping Configuration No. HTFA ASY-0001. Affected: All unexpired lots when used with specific BACT/ALERT® Blood Culture Bottles. HTFA-ASY 0104, Shipping Configuration No. HTFA ASY-0001
Reason
The CAPA investigation found that frequent power cycling of BIOFIRE FILMARRAY TORCH systems can cause degradation of the power switch due to internal arcing, potentially leading to carbon build-up, deformation, and eventual switch failure. Systems with multiple modules may have an increased risk of failure over time if degradation has already occurred., To address this issue, the manufacturer redesigned the power entry module for all BIOFIRE FILMARRAY TORCH System Bases manufactured from 16 September 2024. For units produced before this date, a fused extension cable will be installed to...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-09-25
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/79a2bb66-b5ae-453d-8ba9-61eb3a17de67