Philips Medical Systems Nederland B.v.: Interventional Fluoroscopic X-ray System

Recall Summary Philips is recalling its Azurion S interventional fluoroscopic X-ray systems (Model 722285) because some units may contain solid-state drives that don't meet specifications and could fail over time. These drive failures may cause imaging to stop working, displays to go blank, or the system to lose access to stored or new images. The recall affects systems distributed worldwide, including throughout the United States.

Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product
Philips Azurion S. Model Number: 722285.
Reason
Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-10-06
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands,...

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/78d4cd57-7063-4cdb-a54a-6c2fead80a7f