Neoss 51240 Neoss® Guide - Implant Inserter, Np, Narrow

Recalling firm
Neoss Australia Pty Ltd (NZ)
Product
Neoss 51240 Neoss® Guide - Implant Inserter, NP, Narrow. [Device]. Model: Catalogue number: 51240; Batch 200674 - date of manufacture: 2026-07-27
Reason
The has identified that the guide inserters were produced with incorrect laser marking. The physical product is showing 'S SP'. The correct platform is identified on the product label: 'N NP'. The defect is limited to laser marking only and does not affect the functional characteristics of the device, safety or performance of the instrument if used as intended. The marking is used for device identification only. Based on the technical assessment (CAD/CAM simulation), an NP inserter cannot engage with an SP implant and will rotate in the SP implant., The incorrect laser marking could...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-08-24
Status
Ongoing
Distribution
New Zealand - Healthcare Professional

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/77d4b6e2-2614-4721-bb9f-b85fa41aac0d