Neoss 51240 Neoss® Guide - Implant Inserter, Np, Narrow
- Recalling firm
- Neoss Australia Pty Ltd (NZ)
- Product
- Neoss 51240 Neoss® Guide - Implant Inserter, NP, Narrow. [Device]. Model: Catalogue number: 51240; Batch 200674 - date of manufacture: 2026-07-27
- Reason
- The has identified that the guide inserters were produced with incorrect laser marking. The physical product is showing 'S SP'. The correct platform is identified on the product label: 'N NP'. The defect is limited to laser marking only and does not affect the functional characteristics of the device, safety or performance of the instrument if used as intended. The marking is used for device identification only. Based on the technical assessment (CAD/CAM simulation), an NP inserter cannot engage with an SP implant and will rotate in the SP implant., The incorrect laser marking could...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-08-24
- Status
- Ongoing
- Distribution
- New Zealand - Healthcare Professional
Official recall notice from MEDSAFE
Recall details on Nibble Recalls: https://trynibble.app/recalls/77d4b6e2-2614-4721-bb9f-b85fa41aac0d