Mako Sawing Power Tool Attachment, Reciprocating
Recall Summary Stryker New Zealand Ltd is recalling all lots of the MAKO MICS3 Angled Sagittal Saw Attachment (model 210490), a power tool attachment used in hospitals and medical facilities. The external screws that hold the cover in place may loosen and detach during use, potentially creating a foreign object hazard. This recall affects all units distributed to hospitals, pharmacies, and laboratories in New Zealand.
- Recalling firm
- Stryker New Zealand Ltd
- Product
- Mako Sawing power tool attachment, reciprocating. [Device]. Model: MICS3 Angled Sagittal Saw Attachment - catalogue number - 210490. Affected: All lots
- Reason
- The manufacturer has identified a potential problem with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment. There is potential for the screws to loosen and fall out of the attachment during use, which may lead to a foreign object in patient.
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-03-18
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE