Mako Sawing Power Tool Attachment, Reciprocating

Recall Summary Stryker New Zealand Ltd is recalling all lots of the MAKO MICS3 Angled Sagittal Saw Attachment (model 210490), a power tool attachment used in hospitals and medical facilities. The external screws that hold the cover in place may loosen and detach during use, potentially creating a foreign object hazard. This recall affects all units distributed to hospitals, pharmacies, and laboratories in New Zealand.

Recalling firm
Stryker New Zealand Ltd
Product
Mako Sawing power tool attachment, reciprocating. [Device]. Model: MICS3 Angled Sagittal Saw Attachment - catalogue number - 210490. Affected: All lots
Reason
The manufacturer has identified a potential problem with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment. There is potential for the screws to loosen and fall out of the attachment during use, which may lead to a foreign object in patient.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-03-18
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/7761b590-89d0-4554-8537-29c8eb8c267c