Ventec Life Systems, Inc.: Ventilator, Continuous, Facility Use

Recall Summary Ventec Life Systems is recalling the VOCSN+Pro ventilator (Model PRT-00490-100) distributed nationwide because an internal cable connector may disconnect, causing the device to restart and stop delivering ventilation therapy. This affects ventilators used in medical facilities across the United States and Puerto Rico. The company classified this as a Class II recall, indicating a potential serious health risk.

Recalling firm
Ventec Life Systems, Inc.
Product
VOCSN+Pro (Japan), Model Number: PRT-00490-100, Package PN:PRT-00853-100. used for ventilatory support.
Reason
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-09-17
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Distribution US nationwide including Puerto Rico.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/770e5f69-675f-4369-872d-3750e7be66f0